HF Series Generators
Maintenance
MA-1004R4
1
SECTION 1 INTRODUCTION
The purpose of this Periodic Maintenance is to assure continued safe
performance of the X-ray Generator, to increase serviceability, to reduce the
costs (down time, repairs, etc.) and to assure the safety (personal risk).
The following checks and maintenance procedures, together with the
suggested intervals, are the manufacturer’s recommendation for the most
effective Periodic Maintenance schedule for this Generator.
Service tasks here described must be performed exclusively by service
personnel specifically trained on medical X-ray Generators.
The first Periodic Maintenance Service should be performed six (6) months
after installation, and the subsequent services every twelve (12) months.
Periodic Maintenance Service depends on the working load of the Generator
and X-ray Tube.
Take note in the Data Book all the periodic maintenance services
carried out and the data changes made during any maintenance
service.
Note .