HF Series Generators
Maintenance
MA-1004R4
10
2.6 HV TRANSFORMER CONDITION
The HV Transformer contains “Shell Diala AX” oil.
Check that there is not oil leakage. If found, remove the oil fill plug from the top
of the HV Transformer and verify that the oil level is within 20 mm ( 3/4”) of the
top surface of the HV Transformer. If necessary add oil “Shell Diala AX”.
This point does not apply to the hermetic H V Transfromers (black
aluminium HV Transformers).
Make sure that:
• HV oil in the HV Cable terminals is clean and shows no evidence of
arcing.
• HV Cable terminal rings are tight.
2.7 X-RAY TUBE CONDITION
Make sure that:
• All parts are mechanically secure with no oil leaks.
• HV grease on the HV Cable terminals is clean and shows no evidence
of arcing.
• HV Cable terminal rings are tight.
Note .