HF Series Generators
Calibration
CA-1036R2
1
SECTION 1 INTRODUCTION
This Calibration document provides information and procedures to perform all
the adjustments required to establish an optimal performance of this Generator.
Calibrate the Generator immediately after Configuration is
completed.
DO NOT SUPPLY THE MAIN POWER UNTIL SPECIFICALLY
INSTRUCTED TO DO SO IN THIS DOCUMENT.
THE MAIN CAPACITORS OF THE HIGH VOLTAGE INVERTER
RETAIN A LARGE PORTION OF THEIR CHARGE FOR
APPROX. 3 MINUTES AFTER THE UNIT IS TURNED OFF.
Calibration data is entered in digital form and stored in a non-volatile memory
chip (U3-EEPROM) located on the HT Controller Board, thus no battery
back-up is required.
Calibration procedures must be
performed in the order listed in this
document. Perform only the sections required to calibrate this
unit.