HF Series Generators
Installation
IN-1052R0
1
SECTION 1 INTRODUCTION
The Installation process depends on the Generator and System configuration.
Installation must be performed in the order indicated along this document.
Perform only the sections required to install this Generator.
1.1 TOOLS AND TEST EQUIPMENT
The following hand tools and products are required for the Installation:
• Standard service engineers tool kit.
• Electric drill motor and assorted bits.
• Silicone Insulating Grease (proofing compound).
• Alcohol cleaning agent.
The following test equipment is required for Configuration and Calibration:
• Digital Multimeter.
• Non-invasive kVp Meter.
• Digital mAs Meter.
• Calculator.
• Only for AEC purposes:
G Sensitometer (only for Film).
G Densitometer (only for Film).
G Copper Plates for the Collimator Filter Holder (recommended):
2 units of 1 mm thickness,
1 unit of 0.5 mm thickness,
2 units of 0.2 mm thickness,
1 unit of 0.1 mm thickness.
G Acrylic Plastic Plates can be used Instead of Copper Plates:
6 units of 5 cm. thickness,
5 units of 1cm. thickness.
G Dosimeter (optionally).
• Only for Tomo purposes:
G Tomophantom tool.