HF Series Generators
Maintenance
MA-1004R4
3
SECTION 2 PERIODIC MAINTENANCE PROCEDURES
When any major component, such as a X-ray Tube,
HV Transformer or major circuit board, is replaced in the
system, perform the respective Configuration and
Calibration procedures.
Update and take note in the Data Book any new data entered
in memory.
If the HT Control Board or the ATP Console CPU Board is
replaced, check specially that Extended Memory data have
not been lost or modified with the Board change. Compare
Extended Memory data w ith the values noted in the Data
Book.
Also, make some exposures using different techniques and
Focal Spot and check that mA stations are calibrated
correctly, if not perform Calibration procedures.
Before starting the periodic maintenance procedures, it is recommended to
make a test exposure using the same operating factors and conditions as a
typical exposure.
Perform the X-ray tube warm-up procedure if the tube has not been in use for
approximately one hour. ( Refer to Operator Manual of the C onsole).
2.1 TEST EQUIPMENT
The tools and test equipment required to perform the Periodic Maintenance
Servicearethesameasspecifiedin“Installation” document.