HF Series Generators
Configuration
CF-1018R3
18
Table 2-1
Extended Memory Locations
MEMORY LOCATION FUNCTION VALUE
E01 TUBE-1 -- RAD filament stand-by (Autocalibrated. Not field changeable)
E02 TUBE-1 -- RAD tube type
E03 Low Digital mA Loop Closed (from 10 mA to 80 mA)
E04 AEC-1 calibration
E05 High Digital mA Loop Closed (from 100 mA)
E06 kV Loop
E07 Maximum kW (Factory set. Only field changeable to lower value)
E08 AEC-1 tracking
E09 AEC-2 calibration
E10 AEC-2 tracking
E11 Not used.
E12 AEC Density Scale
E13 TUBE-1 -- Exposure Time adjustment - Delay
E14 Not used.
E15 TUBE-1 -- Exposure Time adjustment - Ceq kV
E16 Not used.
TUBE-2 -- RAD filament stand-by (Autocalibrated. Not field changeable)
E17
TUBE-2 -- FLUORO filament setting
E18 TUBE-2 -- RAD or FLUORO tube type
E19 Maximum FLUORO kV
E20 AEC-3 calibration
E21 Not used.
E22 Not used.
E23 AEC-4 calibration / Photomultiplier AEC calibration (SF camera)
E24 AEC-3 tracking / AEC-4 tracking (equal value for both)
E25 FLUORO mA display calibration at 50 kV
E26 FLUORO mA display calibration at 80 kV
E27 FLUORO mA display calibration at 120 kV
E28 Not used.
E29 TUBE-2 -- Exposure Time adjustment - Delay
E30 Not used.
E31 TUBE-2 -- Exposure Time adjustment - Ceq kV
E32 Not used.
* Note.-- For Fluoroscopic use, value in E17 means “Fluoro filament setting” and must be manually set