HF Series Generators
Configuration
CF-1018R3
28
2. Set dip switches 3000SW2-1 and 3000SW2-2 as indicated below, in
order to enable the selection of times with the Low Speed Starter. Dip
switch3000SW2-1 has t o be switched ONLY with the Generator
powered OFF.
3000SW2-1 (selection enable) 3000SW2-2 (Low Speed Starter)
ON OFF
3. Configure the Rotor and Filament Times by setting the dip switches
3000SW2-4 through 3000SW2-8 per Table 3-1.
Table 3-1
Low Speed: Configuration of Rotor and Filament Times
TUBE-1
3000SW2-7 3000SW2-8
ROTOR ACCELERATION TIME
AND FILAMENT SETTING TIME
OPEN (OFF) CLOSED (ON) OPEN (OFF) CLOSED (ON)
0.8 seconds
- -
1.2 seconds
- -
1.8 seconds
- -
2.7 seconds
- -
TUBE-2
3000SW2-5 3000SW2-6
ROTOR ACCELERATION TIME
AND FILAMENT SETTING TIME
OPEN (OFF) CLOSED (ON) OPEN (OFF) CLOSED (ON)
0.8 seconds
- -
1.2 seconds
- -
1.8 seconds
- -
2.7 seconds
- -
3000SW2-4
FLUORO ROTOR
ND FIL
MENT HOLD-O
ER TIME
OPEN (OFF) CLOSED (ON)
After releasing the Fluoro Pedal, the Rotor stops and the Filament Current goes
back to stand-by.
-
After releasing the Fluoro Pedal, 1 minutes passes before the Rotor stops
and the Filament Current goes back to stand-by.
-
Record the switch c onfiguration in the Data Book.
4. To validate previous configuration, turn the Generator ON, wait until
Error--01 (E01) appears on the Console and turn the Generator OFF.
5. Set dip switch 3000SW2 to the original settings as noted in step-1. (Refer
to Section 1.1.5 for the normal settings of Dip Switch 3000SW2 ).
Note .