c. Once the Hyperterminal is opened on the Touch Screen Console,
configure the connection. For that select on the Menu bar:
File --> Properties --> Connect using: COM4
Configure --> Bits per second: 115200
Data Bits: 8
Parity: None
StopBits: 1
Flow Control: None
d. Then press “OK” and “OK”.
e. Select on the Menu bar: Transfer --> Send File
on this window select:
Filename:
C:\ program files\Rad_Console (this name can be different)
\APR_English.ini (or APR_French.ini, or APR_Spanish.ini,
or APR_German.ini, or APR_Italian.ini, or
APR_Portuguese.ini, according to the APR language
previously selected on the Settings / Settings Menu.)
Use sending protocol:
Xmodem
f. Let that window open without pressing “Send”, by the moment.
4. Press “Reveice” on the Laptop (PC) window. It will ask for the name to
save the file: Type “APR_English.ini” (or APR_French.ini, or
APR_Spanish.ini, or APR_German.ini, or APR_Italian.ini, or
APR_Portuguese.ini, according to the APR language previously
selected on the Settings / Settings Menu.)
5. Press “Send” on the Touch Screen window.
6. W ait until the transference is complete (this can take a few minutes)
7. When the transference is complete, close the Hyperterminal application
in the Laptop and Touch Screen Console. Disconnect the serial cable.
8. On the Touch Screen, press the“Start--Windows” button on the keyboard
and select: “Programs / Start up (select the first one) / Console”, to come
back to the Application.