HF Series Generators
T roubleshooting
TR-1005R4
26
TECHNIQUE ERROR: If it activates during exposure it means that:
The exposure has been interrupted by the “Security Timer”
because of a system failure. Call Field Service. (Also refer
to Error Code “E34”).
This error can also be shown:
• after an APR technique selection to advise that exposure parameters
displayed on the Console are not the values stored for this APR
technique. Exposure parameters are adapted by the Generator to
another enable values. (Also refer to Error Code “E34”).
• after the “ABC” selection, when ABC is not enable. (Also refer to Error
Code “E34”).
• if a failure on the Automatic Collimator has been detected (blades are full
open or in movement during exposure, etc.). In this case the indicator
light will blink. (Also refer to Error Code “E48”).
TECH