HF Series Generators
T roubleshooting
TR-1005R4
41
ERROR CODE : E10
DESCRIPTION : EEPROM corrupted or not initialized in ATP Console CPU Board or in HT Controller Board.
Wrong data calibration.
ERROR TYPE : Fatal Error during power up (when EEPROM U3 in HT Controller Board is corrupted or not
initialized).
Inform ative error during power up (when EEPROM U18 in ATP Console Board is corrupted or not
initialized or when EPROM U24 in ATP Console Board has been changed). In both cases, “E10”
appears together with “E34” Error Code or “Technique Error” indication.
APPLICABLE TO : All Generators.
APPEARS WHEN : Only during initialization phase.
INFORMATION / SYMPTOM : Generator does not continue with start up.
POSIBLE CAUSES
EEPROM U3 in HT Controller Board corrupted or no initialized.
EEPROM U18 in ATP Console Board corrupted or no initialized.
EPROM U24 in ATP Console Board has been changed.
ACTIONS
1. If EPROM U24 in the ATP Console Board has been replaced, reset the Error Code to
acknowledge that the NVRAM has been initialized.
2. If the error does not reset, turn the Generator OFF and set Dip-switch A3024SW2-3 in ON
position to allow Service Mode. Turn the Generator ON and enter in workstations configuration
and check as referred in Service Manual (do not forget to exit from Configuration mode to store
the workstation configuration).
3. If the problem is still present, and “E10” error appears together with “E34” Error Code or
“Technique Error” indication, replace ATP Console Board.
4. If the problem is still present and only “E10” error appears on the Console, replace HT Controller
Board.