HF Series Generators
T roubleshooting
TR-1005R4
49
ERROR CODE : E16
DESCRIPTION : Invalid value of : kVp, mA or kW.
ERROR TYPE : Informative. Does not allow “Prep”or“Ex p”.
APPLICABLE TO : All Generators
APPEARS WHEN : In “Prep”
INFORMATION / SYMPTOM : Selected mA or kVp are not correct.
POSIBLE CAUSES
Wrong maximum kVp configuration.
Wrong software compatibility on the Generator Cabinet and Console.
When pressing “Prep” during calibration of Open Loop mA with a technique that overpasses the
Generator power kVp/mA.
ACTIONS
1. Check the switch 5 of SW2 on HT Controller Board (125 kVp in open position and 150 kVp in
closed position).
2. Check the software version on U5 of HT Controller Board and U24 of ATP Console Board.
3. Set calibration data manually as per calibration procedure table “mA calibration number change”
at calibration section in this manual.
ERROR CODE : E17
DESCRIPTION : Communication error between ATP Console CPU Board and HT Controller Board.
ERROR TYPE : Fatal Error . Generator opens line contactor that remains in a endless loop.
APPLICABLE TO : All Generators
APPEARS WHEN : Once initialization phase is over at any moment.
INFORMATION / SYMPTOM : No communication between Console and Generator Cabinet.
POSIBLE CAUSES
Defective communication cable between Console and Generator Cabinet (J3).
Noise on the bucky circuitry.
Defective HT Controller Board or defective A TP Console Board.
The “Prep” signal from Console to HT Controller is short--circuited to ground.
ACTIONS
1. Turn Generator OFF/ON.
2. If E01 appears, follow procedure for E01.
3. If E01 does not appear, it is an intermittent error due an external device, install a R--C filter in the
power supply and at bucky start circuitry.
4. If error persists, replace HT Controller or ATP Console Boards.