HF Series Generators
T roubleshooting
TR-1005R4
104
ERROR CODE : E93
DESCRIPTION : Incorrect signal measure in ICOM (CH4).
ERROR TYPE : Indicative. Does not allow exposures.
APPLICABLE TO : All Generators with DRAC or L V-DRAC.
APPEARS WHEN : In Stand-by.
INFORMATION / SYMPTOM : E93 is displayed and it is not possible to make exposures.
POSIBLE CAUSES
The system is detecting current through main wire and it should be zero.
ACTIONS
1. Check the voltage at TP1, TP3 and TP36 in Control DRAC Board. It should be 0V.
ERROR CODE : E95
DESCRIPTION : Rapid Termination
ERROR TYPE : Indicative. Does not allow exposures.
APPLICABLE TO : Generators with Rapid Termination application Installed.
APPEARS WHEN : While Exposure. (after 10ms from exposure or after 10% of the exposure back-up time whatever
is first).
INFORMATION / SYMPTOM : E95 is displayed and it is not possible to make Exposures.
POSIBLE CAUSES
The selected Ion Chamber or the selected parameters (short backup time) are not appropriate for an
exposure with AEC
Switch SW1--3 in ATP Console A3024--XX PCB is in ON position.
ACTIONS
1. Press the respective button on the Console to reset the Error indication.
2. Select appropriate Ion Chamber or modify parameters.
3. if error remains check SW1-3 in ATP Console A3024--XX PCB is in OFF position.