HF Series Generators
T roubleshooting
TR-1005R4
106
ERROR CODE : E98
DESCRIPTION : Dip Switch 3024SW2-3 in A TP Console Board set for Configuration and Calibration Mode Active.
ERROR TYPE : Informative. It allows normal operation.
APPLICABLE TO : All Generators.
APPEARS WHEN : After turning on the Control Console.
INFORMATION / SYMPTOM : E98 appears at console display.
POSIBLE CAUSES
Dip Switch 3024SW2-3 in ATP Console Board set in ON” position.
ACTIONS
1. Reset the error condition by pressing the respective button on the Console.
Keep in mind that this error will appear each time the Generator is turned OFF/ON during service
procedures (configuration, calibration, etc.) whenever Dip Switch 3024SW2-3 in A TP Console
Boardisin“ON (closed) position (for service mode allowed).
2. When servicing is finished and the Generator is ready for normal operation, turn the Generator
OFF and set Dip Switch 3024SW2-3 in ATP Console Board in OFF (open) position (operation
mode).