IMPORTANT NOTE FOR AUTO-CALIBRATION
Auto-calibration procedure of “Digital mA Loop Open” (Filament Current Numbers)
is the recommended method to achieve the most accurate calibration. In any case,
Auto-Calibration is a help for the Manual Calibration procedure that can be
performed at any moment or when Auto-calibration procedure is not fully performed.
(Refer to Section 2.5 “Digital mA Loop Open” in the Calibration document).
During the Auto-calibration process, the Generator informs about different conditions
through messages or codes displayed on the Console. The table below is a resume of
these situations:
MESSAGE / CODE SITUATION
“TUBE OVERLOAD”,
“GENERATOR OVERLOAD”,
ETC
The Generator can not calibrate in this moment the s elected kV / mA combination
(because anode overheated, space charge, generator power limit, etc...). In this
case, the Generator will continue with Auto-calibration of the following available
kV/mA combination for the selected Focal Spot. At the end of the process it will try
to calibrate or calculate the combinations previously uncalibrated.
“CONFIRM” / “DISMISS”
This messa
e / code a
ears at the be
innin
of the
uto-calibration
rocedure to
“222”
confirm or leave the Auto-calibration procedure.
“TUBE TOO HOT”
Exposures are inhibited momentarily because the Heat Units capacity that remain
“111”
is 40% or less. Wait until the
-ray Tube begins to cool and recovers the Heat Units
capacity.
“AUTOCALIBRATION
FAILURE”
Auto-calibration procedure has been cancelled after ten attempts of calibration of
the same kV / mA combination. In this case, calibrate manually the k V / mA
combinations uncalibrated.
“777” or “888”
Code “
” appears instead code “
” when
uto-calibration procedure is
cancelled due to “space charge” during calibration of 40 kV at any mA station for
Small Focal Spot.
“AUTO CALIBRATION OK”
uto-calibration
rocedure has been success
ull
erformed for the selected
“999”
Focal Spot.
Auto-calibration procedure is indepen d en t fo r each Fo cal Spot. If Auto-calibration
has been successful for one of the F o cal Spots and aborted for th e o ther Focal Spot,
it is only required to perform the Manual Calibration of the mA station uncalibrated
for t h e o t h er Focal Spot. Enter in “Manual Calibration” mode, select each kV / mA
combination and identify the combinations that require to be manually calibrated
(these combinations keep the Filament Stand-b y valu e as F ilament Current
Number).