HF Series Generators
Installation
IN-1052R0
34
2. Connect each Ion Chamber cable to J 1 (IC1), J2 (IC2), J3 (IC3) or J5
(IC4) of the “AEC Adaptation Board”. The code for the Ion Chamber cable
supplied by the Generator manufacturer is A3253--01.
Same Ion Chamber types have to be installed in consecutive order
starting at J1 (IC1). In the case of using four Ion Chambers, IC3
and IC4 must have the same film/cassette combination.
The Vacutec, AID or similar Ion Chamber type is directly
connected to the “AEC Adaptation Board”. For Comet Ion
Chambers, cable connection has to be made previously through
the Comet Preamplifier PA-021 and then directly to the “AEC
Adaptation Board”.
For other cases, a cable adapter is required. The Generator
manufacturer has the following cable adapters available: GE
(A3082--01), CGR-BVM (A3081--01), Philips Amplimat
(A3080--01/02 with DB-15 or A6727-01/02 with Phenolic
connector) and MEDYS (A6715--01).
Note .
Note .