9. Read DR16.
10.in DR16, enter a simulation weight in the measuring range of the scale.
11.Write DR16.
12.Start the simulation using the "Weight simulation ON (3)" command.
⇒ The notice "SIMULATION" is shown on the main display and a status bit is set.
⇒ From the start of simulation onward, all configured limits and interfaces refer to the
simulation weight.
13.Stop the simulation using the "Weight simulation OFF (4)" command.
TIA Portal procedure
1. Activate service mode: Command code 1.
2. Read DR03: Command code 2003.
3. Enable the weight simulation [weightSimulationEnabled] in DR03: Selection code 1.
4. Write DR03: Command code 4003.
5. Read DR07: Command code 2007.
6. Assign the "simulation weight" function to the I/O process value 5, 6, 7 or 8
[ioProcessValue5 to ioProcessValue8]: Selection code 1.
7. Write DR07: Command code 4007.
8. Deactivate service mode: Command code 2.
9. Enter the simulation weight for the I/O process value selected in DR07 in the "cyclicIoData
> output" structure in the interface DB.
10.Start simulation mode: Command code 3.
⇒ The notice "SIMULATION" is shown on the main display and a status bit is set.
⇒ From the start of simulation onward, all configured limits and interfaces refer to the
simulation weight.
11.Stop simulation mode: Command code 4.
6.13.2 Simulating dosing
Introduction
When you specify a dosing time, a dosing cycle is simulated completely automatically. When
you specify an emptying time, emptying of the scale is simulated completely automatically.
Example of dosing time
A dosing time of 4.5 ms and a set point of 100 kg has the result that the dosing cycle reaches
a set point of 100 kg in a linear manner within 4.5 ms.
Parameter assignment/addressing
6.13 DR16 Configuration of weight simulation
Technology module TM SIWAREX WP351 HF
82 Operating Instructions, 01/2020, A5E47521010-AA