User Manual For - CONTROLLER/DATA RECORDER MultiCon CMC-99/141
Most important parameters of UI modules contd.:
UI12 U24 I24
Number of inputs 12xU + 12xI 24xU 24xI
Hardware limitations
voltage inputs
current inputs
0V ÷ 12V
0mA ÷ 24mA
0V ÷ 12V
-
-
0mA ÷ 24mA
Hardware resolution
voltage inputs
current inputs
1mV
1µA
1mV
-
-
1µA
Precision
voltage inputs
current inputs
0,1% @ 25ËšC
0,1% @ 25ËšC
0,1% @ 25ËšC
-
-
0,1% @ 25ËšC
Covered
MultiCon CMC-99/141
measurement ranges *
type max. type max. type max.
0÷5V
1÷5V
0÷10V
2÷10V
0÷20mA
4÷20mA
0÷6V
0÷6V
0÷12V
0÷12V
0÷22mA
4÷22mA
0÷5V
1÷5V
0÷10V
2÷10V
-
-
0÷6V
0÷6V
0÷12V
0÷12V
-
-
-
-
-
-
0÷20mA
4÷20mA
-
-
-
-
0÷22mA
4÷22mA
Temperature limitations
50 ppm/
Ëš
C 50 ppm/
Ëš
C 50 ppm/
Ëš
C
Internal impedance
voltage inputs
current inputs
50kΩ
type 100Ω
50kΩ
-
-
type 100Ω
Protection
voltage inputs
current inputs
no
50mA auto-reset fuse
no
-
-
50mA auto-reset fuse
Sampling period 10ms 10ms 10ms
Isolation strength 1 min. @ 500V AC 1 min. @ 500V AC 1 min. @ 500V AC
Weight 62 g 68 g 64 g
*
Measurement ranges are limited by software upon hardware inputs ability, check current list of
MultiCon CMC-99/141
measurement ranges at producer's website
Tab. 8.3 Technical data for the module UI12, U24, I24
238