96 Service Manual
7 Settings Import/Export
7.3 Application Presets
The application preset file is in a format of probedataX.X.X.X-oo.tar.gz, X.X.X.X stands for the
current software version, oo stands for the system settings version. X.X.X.X is only for reference,
but oo must be consistent with the system settings version. Otherwise, the application presets could
not be imported.
7.3.1 Customizing Application Presets
Perform the steps as follows,
1. Select an exam type to be applied.
2. Select an imaging mode, and then make settings of imaging parameters.
3. Press the S key on the key panel, and press the Y key to confirm.
Or, press the N key to ignore the setting.
Going to
User Pre-Set Menu
Are you sure(Y/N)
4. Click Create New Exam on the User preset menu, and press the Set key to confirm.
5. Enter the name of the application preset, and click Create Exam.
6. Click Exit, and the application preset you customized displays on the touch screen.
7.3.2 Exporting Application Presets
Perform the steps as follows,
1. Connect the USB drive to the ultrasound system.
2. Choose Menu -> System Settings -> Load Default -> Store -> Copy user setting to USB.
7.3.3 Importing Application Presets
Perform the steps as follows,
1. Connect the USB drive to the ultrasound system.
2. Choose Menu -> System Settings -> Load Default -> Store -> Copy user setting to USB.
NOTE:
Ensure only one
userdataX.X.X.X.-oo.tar.gz
file is in the USB drive. Otherwise, a failure
import may result or the imported file cannot be recognized.
The default system icons would not be replaced or deleted when you import application
presets to the ultrasound system.