128 Service Manual
11 System Maintenance
Table 12-2 Recommended Method for High-level Disinfection
Disinfectant Manufacturer
Active
Ingredient
Level for Active
Ingredient
Contact
Type
Contact
Period
Cidex J&J Glutaraldehyde 2.4% Soak 45-50min
ResertXLHLD STERIS
H y d r o g e n
peroxide
2.0% Soak 8min
To sterilize the intraoperative probe
You should sterilize the intraoperative probe after each use. Perform the steps as follows:
1. Clean the probe.
For details, refer to Section 12.1.3 Cleaning the Probe.
2. Sterilize the intraoperative probe as shown in Table 12-3.
Table 12-3 Recommended Method for Sterilization
Disinfectant Manufacturer
Active
Ingredient
Level for Active
Ingredient
Contact
Type
Contact
Period
Cidex J&J Glutaraldehyde 2.4% Soak 10h
NOTE:
If you do not use the recommended sterilant, make up it as the manufacturer’s instructions.
3. Take out the probe from the sterilant when the recommended soak period expires.
4. Rinse the probe with the fresh running water, and then dry it with a soft dry cloth.
11.2.3 Storage and Transportation
Store and transport the probe to ensure the probe is in a good condition.
To transport the probe
1. Ensure the probe is cleaned and disinfected before the transportation.
2. Place the probe fully into the carrying case, and carefully wind the cable.
3. Ensure no parts of the probe left outside the carrying case before the package.
4. Pack the carrying case with sponge, and place it in a carton.
To store the probe
The following items are recommended to be performed for the storage.
Always store the probe in the probe holder or the specic package.
Ensure the probe holder or the specic package is clean.
Do not expose it in direct sunlight or environmental temperature of sudden changes.