EasyManua.ls Logo

Spectroquant Pharo 300 - Page 109

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectroquant
®
Pharo 300 Operation
109
ba75703d07 04/2014
Create all required procedure vari-
ables.
5 Create a procedure variable
required for the formula with [Add]
and enter a designation, e.g. the
measured parameter.
or
Using [Next], accept all entries on
the page and switch to the next
page.
6 Add another procedure variable
with [Add].
or
Delete a highlighted procedure
variable with [Delete].
7 Using [Next], accept all entries on
the page and switch to the next
page.
Enter the formula.
8 Enter any numbers with <0...9>.
Use [Operators] , <><>
<><> and <START ENTER>
to enter an operator, a function or
a constant.
Use [Variables], <><>
<><> and <START ENTER>
to select a variable.
The formula is displayed after
each step.
Using <> you can delete the last
element of the formula.
Use [Back] to quit the formula edi-
tor.
Procedure variables
16.04.07 9:52
Procedure variables are variables whose current numerical
values have to be entered during the course of the
measurement (e.g. weighted sample or dilution). If a
procedure variable is required to calculate the result: Create a
procedure variable (K) with <Add>.
Back Add Next
Procedure variables
16.04.07 9:52
K 1 V (extract)
K 2 V (sample)
Back Add Delete D Next
Formula entry
16.04.07 9:52
Use the <Operators> softkey to select an operation, function
or constant (e.g.: +, -, *, tan, log, e, Pi). Use the <Variables>
softkey to select an absorbance at a certain wavelength or a
procedure variable. Enter numerals via the keyboard. You can
erase the last entry with <?>.
Back Operators Variables Next

Table of Contents