Spectroquant
®
Pharo 300 Operation
171
ba75703d07 04/2014
Carrying out the
AQA2 check for a
method
Test record A test record is displayed after the check. It can be printed and stored as a
file (in the internal DataB folder or USB memory device at the USB-A connec-
tion, see section 4.11.1).
Sample printout:
Note
Later you can view the last AQA2 test records for all methods monitored with
AQA2 under AQA2 info.
1 Carry out the check like a normal
measurement (see section 4.5.1
to 4.5.3).
2 Insert the cell or
start measurement with
<START ENTER>.
After the measurement is com-
pleted, the result and its evalua-
tion are displayed.
If the check failed, it is possible to
repeat the measurement.
If the check was successful, the
AQA2 check function is finished.
<HOME>
[AQA]
AQA2 check
AQA2 check
16.04.07 9:52
Target value 2.00
To start measurement, insert cell or press
<START/ENTER>
51: 14558
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Pharo 30009130512 1.30-Merck-1.60 Administrator
AQA2 OK
Protocol ID 32
Executed by: Administrator
Executed 21.05.2007
Valid until: 13.08.2007
Method 55: 14543 PO4-P
Standard ID CC10 OC557775
Target value 0.80 +- 0.08 mg/l
Measured value 0.84 mg/l