Spectroquant
®
Pharo 300 Operation
177
ba75703d07 04/2014
Test record The result of the MatrixCheck is displayed in a test record. You can print this
record and save it as a file.
To save the file in the photometer, select the Internal DataB folder as the
location. To save the file in an external USB memory device at the USB-A
connection, select USB memory as the location (see section 4.11.1).
Sample printout:
The display returns to the mea-
sured value display of the original
sample without spiking / dilution.
The [MC] status indicator is
displayed. A MatrixCheck was
carried out for this measured
value.
Concentration
16.04.07 9:52
[MC]
45
mg/l
14: 14540
COD
16 mm
10 - 150 mg/l
Setup
Method list
Citation form Unit
Pharo 30009130512 1.30-Merck-1.60 Administrator
MatrixCheck OK
Protocol ID 7
Method 14: 14540 COD
Sample concentration45 mg/lCOD
Standard ID COD 1500
Standard concentration400 mg/lCOD
Sample Standard Target value
Actual value
ml ml mg/lmg/l
10 0.5 625894% OK
10 1 777192% OK