Spectroquant
®
Pharo 300 What to do if ...
197
ba75703e07 04/2014
Self-test does not
start.
The instrument
displays
Please remove cell
Obviously incorrect
measured values
Fluctuating
measured values
Cause Remedy
A cell is inserted in one of the cell
shafts
Remove the cell
Then press the
<START ENTER> key
A foreign object is inserted in one
of the cell shafts
Remove foreign object
Then press the
<START ENTER> key
The instrument has to carry out a
new adjustment for the rectangu-
lar cell recognition
Press the <START ENTER>
key.
The cell shaft is contaminated Clean the cell shaft (see section
5.2.2 and section 6.1)
Restart the instrument
If necessary, confirm the Please
remove cell message with
<START ENTER>.
Instrument defective Return instrument to service
department
Cause Remedy
Cell contaminated Clean the cell
Dilution set incorrectly Set the dilution
Selected method not suitable Select different method
Zero measurement incorrect Perform zero measurement
Blank value incorrect Remeasure the blank value
Cells that are not recognized (e.g.
some plastic cells) disturb the
AutoCheck measurement and, as
a consequence, falsify measured
values until the next AutoCheck is
carried out
Use suitable cells (see section
7.1 and section 8.1)
Cause Remedy
Cell shaft cover open Close the cell shaft cover