What to do if ... Spectroquant
®
Pharo 300
198
ba75703e07 04/2014
Self test failed.
Connected printer
does not print
Cause Remedy
System test: Instrument defective Return instrument to service
department
Filter test: Instrument defective Return instrument to service
department
Wavelength calibration:
Foreign particle in the cell shaft
Lens smudged
Instrument defective
Remove foreign object
Clean the lens (see section 5.2.3
or section 6.1). If this happens
repeatedly, check the operating
conditions (see section 3.2)
Return instrument to service
department
Cause Remedy
Printer not suitable Connect a printer that can inter-
pret the printer control language
PCL-3