Check the pH of the
sample, specified range
:
pH 3 – 10.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Add 1 level blue
microspoon of Al-1K,
close with the screw cap.
Add 0.25 ml of Al-2K
with pipette, close with
the screw cap, and mix.
Reaction time:
5 minutes
100594
Cell Test
Aluminium
Pipette 6.0 ml of the
sample into a reaction
cell, close with the
screw cap, and mix.
Shake the cell vigorous-
ly to dissolve the solid
substance.
Measuring 0.02 – 0.50 mg/l Al
range: Expression of results also possible in mmol/l.
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-f
or-use
aluminium standard solution Certipur
®
, Cat.No. 119770,
concentration 1000 mg/l Al can be used after diluting
accordingly.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300