Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) w
e recommended
to use Spectroquant
®
CombiCheck 40, Cat.No. 114692.
Ready-for-use aluminium standard solution
Certipur
®
,
Cat.No. 119770,
concentration 1000 mg/l Al, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 40) is highly recommended.
Pipette 5.0 ml of the
sample into a test tube.
Add 1 level blue
microspoon of Al-1
to the test tube and
dissolve the solid
substance.
Add 1.2 ml of Al-2
with pipette and mix.
Measuring 0.10 – 1.20 mg/l Al 10-mm cell
range: 0.05 – 0.60 mg/l Al 20-mm cell
0.020 – 0.200 mg/l Al 50-mm cell
Expression of results also possible in mmol/l.
Add 0.25 ml of Al-3
with pipette and mix.
Reaction time:
2 minutes
Transfer the solution into
a corresponding cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment
.
Aluminium
114825
Test
Check the pH of the
sample, specified range
:
pH 3 – 10.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Important:
To measure in the 50-mm cell, the sample volume and the
volume of the reagents hav
e to be doubled for each.
Alternatively, the semi-microcell, Cat.No. 173502, can be
used.