EasyManua.ls Logo

Spectroquant Pharo 300 - Page 238

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
114739
Cell Test
Measuring 0.010 – 2.000 mg/l NH
4
-N
range: 0.01 – 2.58 mg/l NH
4
0.010 – 2.000 mg/l NH
3
-N
0.01 – 2.43 mg/l NH
3
Expression of results also possible in mmol/l.
Ammonium
Pipette 5.0 ml of the
sample into a reaction
cell
close with the screw
cap,
and mix.
Add 1 dose of NH
4
-1K
using the blue dose-
metering cap, close the
cell with the screw cap.
Shake the cell vigorous-
ly to dissolve the solid
substance.
Reaction time:
15 minutes
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Important:
Very high ammonium concentrations in the sample
produce turquoise-colored solutions (measurement solu-
tion should be yellow-g
reen to green) and false-low read-
ings are yielded. In such cases the sample must be di-
luted (plausibility check).
Check the pH of the
sample, specified range
:
pH 4 – 13.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 50, Cat.No. 114695, or the
Standard solution for photometric applications, CRM,
Cat.No. 125022 and 125023.
Ready-for-use ammonium standard solution
Certipur
®
,
Cat.No. 119812,
concentration 1000 mg/l NH
4
+
, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 50) is highly recommended.

Table of Contents