114559
Cell Test
Important:
Very high ammonium concentrations in the sample
produce turquoise-colored solutions (measurement solu-
tion should be yellow-g
reen to green) and false-low read-
ings are yielded. In such cases the sample must be di-
luted (plausibility check).
Measuring 4.0 80.0 mg/l NH
4
-N
range: 5.2 103.0 mg/l NH
4
4.0 80.0 mg/l NH
3
-N
4.9 97.3 mg/l NH
3
Expression of results also possible in mmol/l.
Pipette 0.10 ml of the
sample into a reaction
cell
close with the screw
cap,
and mix.
Add 1 dose of NH
4
-1K
using the blue dose-
metering cap, close the
cell with the screw cap.
Shake the cell vigorous-
ly to dissolve the solid
substance.
Reaction time:
15 minutes
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Ammonium
Check the pH of the
sample, specified range
:
pH 4 – 13.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 70, Cat.No. 114689, or the
Standard solution for photometric applications, CRM,
Cat.No. 125025, 125026, and 125027.
Ready-for-use ammonium standard solution
Certipur
®
,
Cat.No. 119812,
concentration 1000 mg/l NH
4
+
, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 70) is highly recommended.