EasyManua.ls Logo

Spectroquant Pharo 300 - Page 244

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Antimony in water and
wastewater
Application
Reaction time:
2 minutes
Pipette 4.0 ml of the
sample into an empty
round cell (Empty cells,
Cat.No. 114724).
Add approx. 1.5 g of alu-
minium chloride hexa-
hydrate extra pure
(Cat.No. 101084), close
the cell with the screw
cap.
Shake the cell vigorous-
ly to dissolve the solid
substance.
Add 1.0 ml phosphoric
acid 85 % GR (Cat.No.
100573) with pipette,
close the cell with the
screw cap, and mix.
Add 2 drops of
reagent 1, close the
cell with the screw cap,
and mix.
Reaction time:
3 minutes
Add 2 drops of
reagent 2, close the
cell with the screw cap,
and mix.
Add 2 drops of
reagent 3, close the
cell with the screw cap,
and mix.
Add 5.0 ml toluene GR
(Cat.No. 108325) with
pipette, close the cell
with the screw cap.
Important:
The exact composition and preparation of the reagents 1,
2, and 3 used are given in the corresponding application,
which also includes further inf
ormation on the method
employed. This application can be downloaded directly at
www.analytical-test-kits.com.
Note:
Empty cells with screw caps, Cat.No. 114724 are
recommended for the preparation.
These cells can be
sealed with the screw caps, thus enabling a hazard-free
mixing of the sample.
Measuring range: 0.10 – 8.00 mg/l Sb 10-mm cell
Place the cell into the
cell compartment.
Select method no. 130.
Shake the cell vigorous-
ly for 30 seconds.
Leave to stand to allow
phases to separate.
Transfer the solution into
a rectangular cell.
Aspirate the clear upper
phase from the tube with
pipette.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300

Table of Contents