EasyManua.ls Logo

Spectroquant Pharo 300 - Page 250

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Add 6.0 ml of B-6 with
pipette, close with the
screw cap, and mix.
Reaction time:
2 minutes
Transfer the solution
into a cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.
Pipette 5.0 ml of the
sample in to a test tube
with screw cap. (Impor-
tant: Do not use test
tubes made of glass
containing boron!)
Add 1.0 ml of B-1 with
pipette, close with the
screw cap, and mix.
Add 1.5 ml of B-2 with
pipette and close with
the screw cap.
Shake the tube
vigorously for 1 minute.
Aspirate 0.5 ml of the
clear lower phase from
the tube with pipette.
Measuring 0.050 0.800 mg/l B 10-mm cell
range: Expression of results also possible in mmol/l.
Boron
114839
Test
Transfer the extract to a
separate fresh tube.
Add 0.80 ml of B-3 with
pipette, close with the
screw cap, and mix.
Add 4 drops of B-4,
close with the screw
cap, and mix.
Add 15 drops of B-5,
close with the screw
cap, and mix.
Reaction time:
12 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-f
or-use
boron standard solution Certipur
®
, Cat.No. 119500,
concentration 1000 mg/l B can also be used after
diluting accordingly.
Check the pH of the
sample, specified range
:
pH 1 – 13.
Release 06/2014 - Spectroquant
®