Measuring range: 0.003 – 0.120 mg/l BrO
3
50-mm cell
Attention! The measurement is carried out at 550 nm in a 50-mm rectangular cell against a blank, prepared
from distilled water (Water for analysis EMSURE
®
, Cat.No. 116754, is recommended) and the
reagents in an analogous manner.
Bromate in water and
drinking water
Application
Evaporate 200 ml of
sample solution in a
glass beaker almost to
dryness on the hob.
Place the cell into the
cell compartment.
Select method no. 1195.
Filter turbid samples. Transfer the residue to a
20-ml volumetric glass
using a little distilled
water (Water for analysis
EMSURE
®
, Cat.No.
116754, is recommend-
ed).
Pipette 10 ml of the pre-
treated sample into a
test tube.
Add 0.10 ml of
reagent 1 with pipette
and mix.
Add 0.20 ml of
reagent 2 with pipette
and mix.
Add 0.20 ml perchloric
acid 70 - 72 % GR
(Cat.No. 100519) with
pipette and mix.
Reaction time:
30 minutes
Transfer the solution into
a cell.
Important:
The exact composition and preparation of the reagents 1
and 2 used are given in the corresponding application,
which also includes further inf
ormation on the method
employed. This application can be downloaded directly at
www.analytical-test-kits.com.
Make up the contents of
the volumetric flask to
the mark with distilled
water
(Water for analysis
EMSURE
®
, Cat.No.
116754, is recommend-
ed) and mix thoroughly:
pretreated sample.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
20 ml