Check the pH of the
sample, specified range
:
pH 4 – 8.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Pipette 10 ml of the
sample into a test tube.
Add 1 level blue micro-
spoon of Br
2
-1.
Measuring 0.10 10.00 mg/l Br
2
10-mm cell
range: 0.05 5.00 mg/l Br
2
20-mm cell
0.020 2.000 mg/l Br
2
50-mm cell
Expression of results also possible in mmol/l.
Bromine
100605
Test
Reaction time:
1 minute
Transfer the solution into
a corresponding cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.
Important:
Very high bromine concentrations in the sample produce
yellow-colored solutions (measurement solution should be
red) and f
alse-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a freshly prepared
standard solution
can be used (see section “Standard
solutions”).
Shake vigorously to
dissolve the solid
substance.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300