114730
Cell Test
Measuring 5 125 mg/l Cl
range: Expression of results also possible in mmol/l.
Chloride
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) w
e recommended
to use Spectroquant
®
CombiCheck 10 and 20, Cat.No.
114676 and 114675.
Ready-for-use chloride standard solution Certipur
®
,
Cat.No. 119897, concentration 1000 mg/l Cl
, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
Check the pH of the
sample, specified range:
pH 1 12.
If required, add dilute
ammonia solution or
nitric acid drop by drop
to adjust the pH.
Pipette 0.50 ml of
Cl-1K into a reaction
cell, close with the screw
cap, and mix.
Add 1.0 ml of the sample
with pipette
, close with
the screw cap, and mix.