Pipette 5.0 ml of the
sample into a test tube.
Calcium
100049
Test
Reaction time:
5 minutes
Transfer the solution into
a cell.
Place the cell into the
cell compartment.
Select method no. 304.
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) ready-f
or-use
calcium standard solution Certipur
®
, Cat.No. 119778,
concentration 1000 mg/l Ca, can be used after
diluting accordingly.
Add 0.50 ml of Ca-1 with
pipette and mix.
Check the pH of the
sample, specified range:
pH 3 – 9.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Measuring 0.20– 4.00 mg/l Ca 10-mm cell
range: Expression of results also possible in mmol/l.
Add 0.50 ml of Ca-2 with
pipette and mix.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Important:
A separate calibration must be made for each batch.
It is recommended to perform a calibr
ation with a blank
and 5 standard solutions over the entire measuring range.
The calibration should be checked regularly using stan-
dard solutions.