Filter sample solution
through a membrane
filter with 0.45 µm pore
size.
Notes:
Filtered sample =
true color.
Unfiltered sample =
apparent color.
Transfer the solution
into a corresponding
cell.
Place the cell into the
cell compartment, select
method no. 15, 61, or
78.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Measuring 1 250 m
-1
436 nm 10-mm cell Method No. 015 (436)
range: 0.3 125.0 m
-1
436 nm 20-mm cell Method No. 015 (436)
0.1 50.0 m
-1
436 nm 50-mm cell Method No. 015 (436)
1 250 m
-1
525 nm 10-mm cell Method No. 061 (525)
0.3 – 125.0 m
-1
525 nm 20-mm cell Method No. 061 (525)
0.1 50.0 m
-1
525 nm 50-mm cell Method No. 061 (525)
1 250 m
-1
620 nm 10-mm cell Method No. 078 (620)
0.3 125.0 m
-1
620 nm 20-mm cell Method No. 078 (620)
0.1 50.0 m
-1
620 nm 50-mm cell Method No. 078 (620)
Color
(Spectral Absorption Coefficient)
analogous to EN ISO 7887