Swirl the cell to dissolve
the solid substance.
Add 5.0 ml of distilled
water (Water for analy-
sis EMSURE
®
, Cat.No.
116754, is recommend-
ed) with pipette, close
with the screw cap, and
mix.
Measuring 2 – 160 mg/l cyanuric acid 20-mm cell
range: Expression of results also possible in mmol/l.
Cyanuric Acid
119253
Test
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) a cy
anuric acid standard
solution must be prepared from Cyanuric acid, Cat.No.
820358 (see section “Standard solutions”).
Transfer the solution into
a rectangular cell.
Select method with
AutoSelector.
Place the cell into the
cell compartment.
Filter turbid samples.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Add 1 reagent tablet
Cyanuric Acid, crush
with stirring rod, and
close with the screw cap.
Pipette 5.0 ml of the
sample into into an
empty test tube (e. g.
flat-bottomed tubes
cells, Cat.No. 114902).