Cell Test
Nitrate
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
Pipette 0.50 ml of the
sample into a reaction
cell, do not mix.
Add 1.0 ml of NO
3
-1K
with pipette
, close the
cell with the screw cap,
and mix. Caution, cell
becomes hot!
Reaction time:
10 minutes
Measuring 1.0 50.0 mg/l NO
3
-N
range: 4 221 mg/l NO
3
Expression of results also possible in mmol/l.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) w
e recommended to use
Spectroquant
®
CombiCheck 80, Cat.No. 114738, or the
Standard solution for photometric applications, CRM,
Cat.No. 125037, 125038, and 125039.
Ready-for-use nitrate standard solution Certipur
®
,
Cat.No. 119811, concentration 1000 mg/l NO
3
, can also
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 80) is highly recommended.