Spectral Attenuation Coefficient
(254)
analogous to DIN 38404
Shake the unfiltered
sample solution to
e
venly suspend the
turbidity-causing sub-
stances
. Do not disperse
the contents, measure
immediately.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Note:
When the turbidity correction function is activated (see Descrip-
tion of Function, section 4.5.9 “Automatic Turbidity correction”),
the corrected spectral attenuation coefficient (254)korr can
be determined.
The turbidity correction is carried out as per DIN 38404 at
550 nm.
Measuring 3 – 250 m
-1
254 nm 10-mm cell
range: 1 – 125 m
-1
254 nm 20-mm cell
0.5 – 50.0 m
-1
254 nm 50-mm cell
Transfer the solution
into a corresponding
cell.
Place the cell into the
cell compartment, select
method no. 301.