Transfer the solution
into a corresponding
cell.
Place the cell into the
cell compartment, select
method no. 302.
Measuring 3 – 250 m
-1
436 nm 10-mm cell
range: 1 – 125 m
-1
436 nm 20-mm cell
0.5 – 50.0 m
-1
436 nm 50-mm cell
Spectral Absorption Coefficient
(436)
analogous to EN ISO 7887
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300
Filter sample solution
through a membrane
filter with 0.45 µm pore
size.
Notes:
Filtered sample =
true color.
Unfiltered sample =
apparent color.