Pipette 4.0 ml of the
sample into a reaction
cell. Close with the
screw cap.
Shake the cell for
1 minute vigorously.
Swirl the cell before
measurement.
Place the cell into the
cell compartment. Align
the mark on the cell with
that on the photometer.
101787
Cell Test
Surfactants (nonionic)
Measuring 0.010 – 7.50 mg/l surfactants (nonionic)
range: (calculated as Triton
®
X-100)
Reaction time:
2 minutes
Quality assurance:
To check the measurement system (test reagents,
measurement device, and handling) a surf
actants
standard solution must be prepared from Triton
®
X-100,
Cat.No. 112298 (see section “Standard solutions”).
Check the pH of the
sample, specified range
:
pH 3 – 9.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Release 06/2014 - Spectroquant
®
UV/VIS Spectrophotometer Pharo 300