Operation Spectroquant
®
Pharo 300
50
ba75703d07 04/2014
Note
The result of the self-test can be viewed and printed with the [Info] function
key (see section 4.18).
Warm-up time After being switched on the photometer requires a warm-up time of 15 min-
utes. Reproducibility of measurement data is restricted during the warm-up
time.
Therefore, do not measure during the warm-up time.
During the warm-up time, a progress bar appears on the display next to the
date. The progress bar disappears as soon as the warm-up time is over.
The self-test includes:
– the test of the memory, proces-
sor,
internal interfaces,
filter and lamp
– a calibration for each wave-
length
After the self-test is completed,
the main menu is displayed.
Self test
16.04.07 9:52
Keep cover closed
System test
Filter test
Lamp test
Wavelength calibration
Progress bar during warm-up
time