Operation Spectroquant
®
Pharo 300
62
ba75703d07 04/2014
For rectangular cells, the zero adjustment must be carried out with the
same cell type (manufacturer and cell material [e.g. optical glass, quartz
glass, plastic]) that is used for measurement. This is important because
cells of different manufacturers have a different absorption behavior.
When changing the cell type repeat the zero adjustment with the new type.
Prior to zero adjustment, clean the rectangular cell and fill it with distilled
water. The minimum filling level is 20 mm.
Rectangular cells always have to be inserted in the cell shaft with the
same orientation for measurement and zero adjustment (e.g. cell printing
on the left side ).
Note
Ordering information is given in chapter 8 A
CCESSORIES AND OPTIONS. The
cells listed in the chapter 8 A
CCESSORIES AND OPTIONS are especially adapted
to the Merck Spectroquant
®
test set program. General requirements of the
cells are given in chapter 7 T
ECHNICAL DATA. Note that the spectral transpar-
ency of the cell must be suitable for the intended application (example, quartz
cell for UV range).