EasyManua.ls Logo

Spectroquant Pharo 300 - Page 63

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectroquant
®
Pharo 300 Operation
63
ba75703d07 04/2014
Carrying out a zero
adjustment
The zero adjustment takes place similarly in the Concentration, Absorbance
/ % Transmission, Special / Multi wavelengths and Kinetics modes.
1 In the respective mode, press the
<BLANK ZERO> key.
2 In Concentration mode only:
Select and confirm Zero adjust-
ment.
The zero adjustment window pops
up.
3 Close the inner turn-up lid.
4 Depending on the cell type, insert
the zero cell as follows:
Round cell:
Insert the round cell in the round
cell shaft so it touches the bottom.
If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.
Concentration
16.04.07 9:52
51: 14558
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Setup
Method list
Citation form Unit
Adjust
Blank value
Zero adjustment
Zero adjustment
16.04.07 9:52
Please insert zero cell (distilled water) or
press <START/ENTER>
Zero cell
(H O dist.)
2
inner
turn-up lid

Table of Contents