Operation Spectroquant
®
Pharo 300
80
ba75703d07 04/2014
User-defined
methods
For user-defined methods, you can activate the reagent blank value function
as follows only:
Note
If no value for E0 is stored during the entry of value pairs or the measurement
and storing of standard solutions for a nonlinear function (parabola or
polygon line), the message, No blank value correction is intended for this
method. appears when the User-defined blank value function is activated.
The blank value (E0) can be entered later by editing the method.
Entry type Function type Reagent
blank value
possible?
Entry of a function
(with and without entering the ordi-
nate intercept)
Linear Yes
Nonlinear No
Entry of value pairs or measure-
ment and storage of standard solu-
tions
(with entering/measuring and stor-
ing E0)
Linear Yes
Parabola
(second-order function)
Yes
Polygon line No
Entry of value pairs or measure-
ment and storage of standard solu-
tions
(without entering/measuring and
storing E0)
Linear Yes
Parabola
(second-order function)
Polygon line
Polygon line through
zero
No