3 Insert the cell with the blank sam-
ple.
The first single measurement for
the reagent blank value takes
place.
The following data is displayed as
the result:
– The measured absorbance
from the (last) single measure-
ment.
– The median from all single mea-
surements carried out up to
now.
4 If necessary, carry out further sin-
gle measurements for the forma-
tion of the median with [Next
meas.]
or
discard the last single measure-
ment with [Discard].
5 To accept the median value, press
[Apply].
The Lot number entry field pops
up.
6 Enter and confirm the Lot number
(<A...9>).
The blank value measurement is
completed.
The photometer is ready to mea-
sure.
The use of the reagent blank value
is indicated by [BV/Lot number] in
the top right corner of the display.
Blank value
16.04.07 9:52
To start measurement, insert cell or press
<START/ENTER>
51: 14558
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Blank value
16.04.07 9:52
Last measured absorbance
0.600
Median
0.600 (1 Measurement(s))
51: 14558
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Next meas.
Discard Apply
Blank value
16.04.07 9:52
[BV/Lot number]
To start measurement, insert cell or press
<START/ENTER>
51: 14558
NH
4
-N
16 mm
0.20 - 8.00 mg/l