Operation Spectroquant
®
Pharo 300
82
ba75703d07 04/2014
3 Insert the cell with the blank sam-
ple.
The first single measurement for
the reagent blank value takes
place.
The following data is displayed as
the result:
The measured absorbance
from the (last) single measure-
ment.
The median from all single mea-
surements carried out up to
now.
4 If necessary, carry out further sin-
gle measurements for the forma-
tion of the median with [Next
meas.]
or
discard the last single measure-
ment with [Discard].
5 To accept the median value, press
[Apply].
The Lot number entry field pops
up.
6 Enter and confirm the Lot number
(<A...9>).
The blank value measurement is
completed.
The photometer is ready to mea-
sure.
The use of the reagent blank value
is indicated by [BV/Lot number] in
the top right corner of the display.