EasyManua.ls Logo

Spectroquant Pharo 300 - Page 89

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectroquant
®
Pharo 300 Operation
89
ba75703d07 04/2014
Measuring with user
calibration
Select the method manually
(see section 4.5.5).
If there are already data for the
zero adjustment, reagent blank
value or a user calibration avail-
able, the photometer informs you
of this. You can take over or dis-
card the available values.
If the available user calibration
should not be used, a query with
further options pops up:
- Use default calibration
The existing user calibration is
erased. Further measurements
will be carried out with the origi-
nal calibration stored with the
method
- Recalibrate
The existing user calibration is
erased. A new user calibration
is started.
- Cancel
The existing user calibration
remains stored. The previous
query is displayed.
The photometer is ready to mea-
sure after all the necessary data
have been confirmed or mea-
sured.
<HOME>
Concentration
Concentration
16.01.12 9:52
[Cal][BV/2c][ZERO 10.01.2012 11:08]
To start measurement, insert cell or press
<START/ENTER>
304: Ca
Ca
10 mm
0..20 - 4.00 mg/l
Setup
Method list
Citation form Unit
User calibration
A calibration dated xxx is available for this
method. Should it be used?
Yes
No
Concentration
16.01.12 9:52
[Cal][BV/2c][ZERO 10.01.2012 11:08]
To start measurement, insert cell or press
<START/ENTER>
304: Ca
Ca
10 mm
0..20 - 4.00 mg/l
Setup
Method list
Citation form Unit

Table of Contents