Operation Spectroquant
®
Pharo 300
94
ba75703d07 04/2014
Further method data
* necessary inputs
** default: mg/l
Note
If a nonlinear calibration curve is programmed for a method, it may occur that
the presetting of the following menu items cannot be changed:
Blank required
Calibration possible
Calibration required
Input field Possible entries
Number* 1001 ... 1100
Designation Any name (max. 18 characters)
Version Any version designation (max. 18 characters)
Wavelength* Freely selectable (in nm)
Cell* 16 (round), 10, 20 or 50 mm
Citation form e.g. PO4-P (max. 18 characters)
Unit** e.g. mg/l (max. 18 characters)
Resolution* 0.001, 0.01, 0.1 or 1
Lower and upper limit of
the measuring range *
Any value between zero and the highest concen-
tration of the used standard solutions
Timer 0 to 3 Up to four analysis timers freely adjustable
AQA2 target value Any value within the measuring range
AQA2 tolerance Any
Required measure-
ments
1 or greater
Number of measurements after which a measured
value is documented. With more than one mea-
surement, the documented measured value is the
median from all measurements.
Blank required Yes / No
Calibration possible Yes / No
Calibration required Yes / No