Spectroquant
®
Pharo 300 Operation
95
ba75703d07 04/2014
How to program
user-defined
methods
Note
During the following proceeding, you can return to the previous page at any
time with [Back], e. g. if you want to correct entries, add further value pairs or
eliminate outliers.
1 Enter the general method data
here. The next available method
number is already entered as the
number.
You have the following options
when filling out the input fields:
- Fill out all empty input fields one
after the other
-Using [Method list], select an
already existing method as a
model, give it a new method
number and adjust the entries
-Using [Method list], select an
existing method in order to
change it (without changing the
number).
- You can delete the method
completely with [Delete].
2 Select the menu item, Calibration
curve. Select the method for the
determination of the calibration
line. The following variants can be
selected:
- Measure standard solutions
- Enter value pairs
- Enter formula
3 Using [Next], accept all entries on
the page and switch to the next
page.
<HOME>
Concentration
[Setup]
New method
Edit method
16.04.07 9:52
Number 1001
Designation Nitrite
Version 01
Wavelength 525
Cell 10 mm
Citation form NO2-N
Unit mg/l
Resolution 0.001
Calibration curve Measure standard solutions
Method list Delete Next