SV 971A User Manual
124
Windscreen
Digital filter that compensates the effect of the
SA 22 windscreen.
Chapter
4.7
Range
Position that enables selecting the linear operating
range for the sinusoidal signal selection: Low or
Normal. The calibration factor is always added to
the range limits.
The ranges depend on the selected
Compensation Filter.
Chapter
4.6
Low
Low linear operating range for the sinusoidal signal.
The calibration factor is always added to the range
limits.
The range depends on the selected
Compensation Filter.
Chapter
4.6
Appendix C
Normal
Normal linear operating range for the sinusoidal
signal.
The calibration factor is always added to the range
limits.
The range depends on the selected
Compensation Filter.
Chapter
4.6
Appendix C
Measurement
Trigger
Screen that enables configuring triggering of the
measurement/integration process with parameters:
Trigger, Source, Level and Gradient.
Chapter
4.2
Trigger
Position that switches Off or on the measurement
trigger by selecting its type: Level+, Level or
Gradient+.
If the instrument is waiting for the trigger condition,
the appropriate trigger icon is flashing on the display
alternatively with the „play” icon.
Chapter
4.2
Slope+
Type of trigger that starts the measurement/
integration by the duration of the Integration Period
on condition: rising value of the RMS result (Source)
integrated during 0,5 ms passes above the
threshold value (Level).
Chapter
4.2
Slope -
Type of trigger that starts the measurement/
integration by the duration of the Integration Period
on condition: falling value of the RMS result
(Source) integrated during 0,5 ms passes below the
threshold value (Level).
Chapter
4.2