SV 971A User Manual Appendixes
142
Note: All bytes of that transmission are ASCII characters.
Note: Changing settings using #1 functions during measurements running state (#1,S1;) is
blocked. Stop the measurements (#1,S0;) before changing the settings.
In order to read out all current control settings the user should send to the device the following sequence of
characters:
#1;
In this case the instrument outputs all control settings given in A.3 Control setting codes in the format:
#1,Xccc,Xccc,(...),Xccc;
Example: The instrument sends the following sequence of characters as an answer for the mentioned above
request:
#1,U971,N1234,W1.05.1,Q0.01,M1,R1,F2:1,F3:2,F1:3,F2:4,F3:5,F1:6,J2:1,J3:2,J1:3,J3:4,J3:5,J1:6,f1,C
1:1,C0:2,C2:3,C1:4,C0:5,C2:6,B0:1,B3:2,B15:3,b0,d1s,D10s,K5,L0,Y3,y0,XT0,XL100,XQ0,Xq0,XC115:
1,XC115:2,XC115:3,Xl115:1,Xl115:2,Xl115:3,XA0,XD-1:1,XD-1:2,XD-1:3,XD-1:4,XD-1:5,XD-
1:6,S0,T1,e480,c1:1,c1:2,c1:3,h0:1,h0:2,h0:3,x3:1,x3:2,x5:3,m0,s0,l100,O10,o0,t0;
means that:
SV 971A is investigated (U971); see #7,US; command for unit subtype information;
its number is 1234 (N1234);
software version number is 1.05.1 (W1.05.1);
calibration factor is equal to 0.01 dB (Q0.01);
Level Meter is selected as the measurement function (M1);
range is Low (R1);
A filter is selected in profile 1, SLM function (F2:1);
C filter is selected in profile 2, SLM function (F3:2);
Z filter is chosen in profile 3, SLM function (F1:3);
A filter is selected in profile 1, SEM function (F2:4);
C filter is selected in profile 2, SEM function (F3:5);
Z filter is chosen in profile 3, SEM function (F1:6);
A Peak filter is selected in profile 1, left channel, SLM function (J2:1);
C Peak filter is selected in profile 2, left channel, SLM function (J3:2);
Z Peak filter is selected in profile 3, left channel, SLM function (J1:3);
C Peak filter is selected in profile 1, both channels, SEM function (J3:4);
C Peak filter is selected in profile 2, both channels, SEM function (J3:5);
Z Peak filter is selected in profile 3, both channels, SEM function (J1:6);
Z filter is selected for 1/1 Octave or 1/3 Octave analysis (f1)
Fast detector is selected in profile 1, SLM function (C1:1);
Impulse detector is chosen in profile 2, SLM function (C0:2);
Slow detector is selected in profile 3, SLM function (C2:3);
Fast detector is selected in profile 1, SEM function (C1:4);
Impulse detector is chosen in profile 2, SEM function (C0:5);
Slow detector is selected in profile 3, SEM function (C2:6);
logger’s buffer is not filled by the results from profile 1 (B0:1);
Lpeak and Lmax values are stored in the files of the logger from profile 2 (B3:2);
Lpeak, Lmax, Lmin and Leq values are stored in the files of the logger from profile 3 (B15:3);
results of 1/1 Octave or 1/3 Octave analysis are not stored in the files of the logger (b0);